Sekisui Diagnostics P.E.I. Inc. recalls Sekisui Diagnostics Ammonia L3K¿ Assay; Catalog Number: 293-80-91 and 293-10 For the IN VITRO quantitative measurement…
- Recall date
- November 29, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1078-2017
- FDA classification
- Class II
- Brand / firm
- Sekisui Diagnostics P.E.I. Inc.
- Sold / distributed
- Worldwide Distribution - US including DE, KS, FL, TX, CA, MI, TN, MA, ID, NJ, RI, IL and Internationally to Austria, Mexico, Canada, India, South Korea, China, Spain, South Korea, and Sri Lanka
Why it was recalled
Certain lots are showing an atypical decrease in optical density (OD) over time, which may result in an impact to performance at the high end of the linear range.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Sekisui Diagnostics Ammonia L3K¿ Assay; Catalog Number: 293-80-91 and 293-10 For the IN VITRO quantitative measurement of ammonia concentration in plasma