Medical device recalls Moderate risk

Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Recall date
May 30, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0087-2023
FDA classification
Class II
Brand / firm
Shanghai United Imaging Healthcare Co., Ltd.
Sold / distributed
Distribution to US states of TX, WI, CA, and MI

Why it was recalled

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.