Shanghai United Imaging Healthcare Co., Ltd. recalls uDR 596i Digital Medical X-Ray Imaging System - Product Usage: intended to use by a qualified/ trained doctor or techni…
- Recall date
- February 27, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2521-2020
- FDA classification
- Class II
- Brand / firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Sold / distributed
- US Nationwide distribution including in the state of Texas.
Why it was recalled
Two issues were identified with the stationary x-ray system. 1. Some position illustrations in the patient exam interface show the incorrect anatomical position for the standard of care. When the image is acquired with the wrong reference position, clinical diagnostic requirements cannot be met.T2. Where image stitching is used, the left and right tag information may be lost. When these problems are encountered, an additional exposure is needed and the patient will receive more radiation.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
uDR 596i Digital Medical X-Ray Imaging System - Product Usage: intended to use by a qualified/ trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, limbs and trunk.