Shanghai United Imaging Healthcare Co. product recalled over laceration hazard
- Recall date
- July 16, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Shanghai United Imaging Healthcare Co., Ltd. recalls uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthc…
- Recall number
- Z-2799-2020
- FDA classification
- Class II
- Brand / firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Sold / distributed
- US Nationwide distribution including in the state of Texas.
Why it was recalled
1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.