Medical device recalls Moderate risk

Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

Recall date
October 9, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0553-2025
FDA classification
Class II
Brand / firm
Shimadzu Corp. Analytical & Measuring Instruments Div.
Sold / distributed
Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.

Why it was recalled

There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL