Autosampler recalled over fire hazard
- Recall date
- October 9, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
- Recall number
- Z-0554-2025
- FDA classification
- Class II
- Brand / firm
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.
Why it was recalled
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Get recall alerts
Free email alert whenever Shimadzu Corp. Analytical & Measuring Instruments Div. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Shimadzu Corp. Analytical & Measuring Instruments Div.