Medical device recalls Moderate risk

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION recalls Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the followi…

Recall date
September 6, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0139-2025
FDA classification
Class II
Brand / firm
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Sold / distributed
US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.

Why it was recalled

If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.

Get recall alerts

Free email alert whenever SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION