Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur¿ HBc IgM

Recall date
January 2, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1690-2018
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Nationwide and Canada, Mexico, and Thailand

Why it was recalled

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

ADVIA Centaur¿ HBc IgM