Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator - product Usage: For in vitro diagno…

Recall date
November 19, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1260-2020
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Worldwide distribution  US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WY and countries of UK, CA, PH, MX, VN, TW, TH, IN, CO, AE, AR, AU, BD, BE, BG, BR, CL…

Why it was recalled

Atellica and ADVIA hsCRP calibrator accuracy has drifted away from the reference materials (ERM-DA470/IFCC), which is causing a high bias on QC and patient sample recovery.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator - product Usage: For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the CardioPhase¿ hsCRP assay.