Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Fructosamine Reagent (FRUC), Siemens Material Number 10361941, UDI Number 00630414558073 - Product Usag…
- Recall date
- May 4, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2280-2020
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Mary…
Why it was recalled
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
ADVIA Chemistry Fructosamine Reagent (FRUC), Siemens Material Number 10361941, UDI Number 00630414558073 - Product Usage: use in the quantitative measurement of glycated protein (fructosamine) in human serum and plasma for monitoring the blood glucose levels of diabetic patients over the preceding two-three weeks.