Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Lipase, Ref. No. 01984894 B01-4840-01, Siemens Material No. 10311896, UDI 00630414482446 - Product Usag…

Recall date
December 20, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1363-2020
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CT, FL, GA, IA, KS, MD, MI, NJ, NM, NY, OH, OK, PA, TN, TX, WA.

Why it was recalled

Firm has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using this product.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

ADVIA Chemistry Lipase, Ref. No. 01984894 B01-4840-01, Siemens Material No. 10311896, UDI 00630414482446 - Product Usage: Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas, such as acute pancreatitis and obstruction of the pancreatic duct.

Get recall alerts

Free email alert whenever Siemens Healthcare Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Healthcare Diagnostics, Inc.