Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink…

Recall date
December 16, 2014
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0987-2015
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Distributed in the state of WA.

Why it was recalled

Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.

Get recall alerts

Free email alert whenever Siemens Healthcare Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Healthcare Diagnostics, Inc.