Siemens Healthcare Diagnostics, Inc. recalls Atellica CH 930 Analyzer, Siemens Material Number 11067000 with software version v1.21.0 SP1 or lower
- Recall date
- February 27, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1797-2020
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- US Nationwide. Global distribution.
Why it was recalled
Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or Auto-Dilution results may generate a result of Zero (0) for quantitative assays or Negative for qualitative assays if a well of the reagent pack is not calibrated.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Atellica CH 930 Analyzer, Siemens Material Number 11067000 with software version v1.21.0 SP1 or lower