Siemens Healthcare Diagnostics product recalled over hepatitis A risk
- Recall date
- February 13, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Total Bilirubin_2 assay-In vitro Diagnostic for the determination of total bilirubin SMN: 11097531 - Produc…
- Recall number
- Z-1537-2020
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Nationwide Foreign: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, Un…
Why it was recalled
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Atellica CH Total Bilirubin_2 assay-In vitro Diagnostic for the determination of total bilirubin SMN: 11097531 - Product Usage: Atellica TBil_2 assay is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates using the Atellica¿ CH Analyzer. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block, A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Get recall alerts
Free email alert whenever Siemens Healthcare Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Siemens Healthcare Diagnostics, Inc.