Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urine Albumin (UAlb). Material Number: 11537225
- Recall date
- January 20, 2026
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1484-2026
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania…
Why it was recalled
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Get recall alerts
Free email alert whenever Siemens Healthcare Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.