Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls Dimension Lipase Flex¿ Reagent Cartridge - ( Lipase-esterase, enzymatic, photometric, lipase)

Recall date
December 14, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1796-2018
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Domestic and Foreign consignees

Why it was recalled

Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-to-lot limits.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Dimension Lipase Flex¿ Reagent Cartridge - ( Lipase-esterase, enzymatic, photometric, lipase)

Get recall alerts

Free email alert whenever Siemens Healthcare Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Healthcare Diagnostics, Inc.