Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista 500, Analyzer, chemistry Model no. 10284473
- Recall date
- December 12, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1014-2019
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Nationwide domestic distribution. Worldwide foreign distribution.
Why it was recalled
When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Dimension Vista 500, Analyzer, chemistry Model no. 10284473