Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista Gentamicin Flex reagent cartridge, DV GENT, gentamicin enzyme immunoassay. Affected lots are 17135BC, 1…
- Recall date
- April 19, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2182-2018
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Nationwide distribution. Foreign distribution to Argentina, Australia, Brazil, Bahamas, Canada, Germany, India, Japan, Myanmar, New Zealand, Saudi Arabia, Taiwan, and South Africa.
Why it was recalled
The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators. The negative bias observed for patient, QC, and calibrator samples, when using any of the listed lots, has the potential to impact the interpretation of gentamicin results during therapeutic monitoring at concentrations up to approximately 2.5 ¿g/mL [5.4 ¿mol/L].
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Dimension Vista Gentamicin Flex reagent cartridge, DV GENT, gentamicin enzyme immunoassay. Affected lots are 17135BC, 17150BF, 17159BD, 17178AC, 17269BD, and 17312BB.