Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista Lipase Flex Reagent Cartridge, Model Number 10461745 Product Usage: Dimension Vista - The LIPL method i…
- Recall date
- September 7, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0466-2019
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Worldwide - US Nationwide Distributions The products were distributed to the following foreign countries: Argentina, Australia, Bangladesh, Brazil, Canada, Chile, China, Colombia, Germany, Ecuador, Honduras, Indonesia, India, Japan, Republic of Korea, Mexico, Malaysia, Nicaragua, New Zealand, Peru,…
Why it was recalled
There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Dimension Vista Lipase Flex Reagent Cartridge, Model Number 10461745 Product Usage: Dimension Vista - The LIPL method is an in vitro diagnostic test for the quantitative measurement of lipase in human serum and plasma on the Dimension Vista System.