Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ MULTI 2 SDIL, KD694, SMN# 10483586

Recall date
March 21, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2438-2018
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
The products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The products were distributed to…

Why it was recalled

The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Dimension Vista¿ MULTI 2 SDIL, KD694, SMN# 10483586