Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista System B2 Microglobulin Flex reagent cartridge and urine stabilizer Catalog #K7024 SMN: 10445889 Lot #1…

Recall date
November 3, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1030-2016
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Worldwide Distribution and US in the state of CA.

Why it was recalled

Siemens Healthcare Diagnostics has confirmed an increase in the rate of Abnormal Assay errors and calibration failures with the Dimension Vista B2MIC Flex reagent cartridge: Dimension Vista System Flex reagent cartridge and urine stabilizer B2MIC. In the unlikely event that a patient result is generated, a suppressed result would be in conflict with the patients clinical status and other markers of renal injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Dimension Vista System B2 Microglobulin Flex reagent cartridge and urine stabilizer Catalog #K7024 SMN: 10445889 Lot #15246MA Product Usage: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of B2-microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista System. Measurements of B2-microglobulin aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Get recall alerts

Free email alert whenever Siemens Healthcare Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Healthcare Diagnostics, Inc.