Siemens Healthcare Diagnostics, Inc. recalls IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 10706293 (US Kit Catalog) Product…
- Recall date
- December 21, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1120-2017
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Worldwide Distribution - US Nationwide and the countries of Denmark, Brazil, Croatia and Germany
Why it was recalled
Negative bias results with the IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit was confirmed for two lots (D0119 and D0120). The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 10706293 (US Kit Catalog) Product Usage: For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.