Medical device recalls Moderate risk

Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This system te…

Recall date
July 28, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2479-2015
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics Inc
Sold / distributed
Worldwide Distribution - US (nationwide) Distribution and to the countries of : Albania, Algeria, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria,Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany,G…

Why it was recalled

When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either: duplicate data, missing data, or data from a different patient.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples.