Siemens Healthcare Diagnostics Inc recalls RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778
- Recall date
- May 27, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2432-2020
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics Inc
- Sold / distributed
- Foreign distribution only.
Why it was recalled
The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778