Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls Syva Emit Methotrexate Assay, is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of meth…

Recall date
March 8, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1618-2016
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Units were shipped to MI, FL, NJ, NY, DC, CA, GA, MI, MT, VA, NC, TX, PA, Australia, Belgium, Czech Republic, Finland, Japan, Republic of Korea, Italy, Saudi Arabia, Slovakia, Slovenia, Spain and Switzerland.

Why it was recalled

Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate Application (DVMTX.1) for Dimension Vista System does not meet On-Board Stability claim of 20 hours. Testing using reagent at the end of the 20 hours On-Board Stability has shown that values of methotrexate for patient and quality control samples increased.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Syva Emit Methotrexate Assay, is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of methotrexate in human serum or plasma.