Siemens Healthcare Diagnostics, Inc. recalls The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ cl…
- Recall date
- May 5, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1217-2017
- FDA classification
- Class III
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Distributed throughout the United States
Why it was recalled
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Enzyme II calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
Class III means the product violates regulations but is not likely to cause harm.
What was recalled
The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ clinical chemistry system. For prescription use only.