Medical device recalls Moderate risk

Siemens Medical Solutions USA, Inc recalls Artis Q biplane, Material no. 10848282, for angiography and whole body radiographic/fluoroscopic procedures.

Recall date
May 1, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2303-2018
FDA classification
Class II
Brand / firm
Siemens Medical Solutions USA, Inc
Sold / distributed
Worldwide distribution. US nationwide including Puerto Rico, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Myanmar, Netherlands, New…

Why it was recalled

After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Artis Q biplane, Material no. 10848282, for angiography and whole body radiographic/fluoroscopic procedures.

Get recall alerts

Free email alert whenever Siemens Medical Solutions USA, Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Medical Solutions USA, Inc