Medical device recalls Moderate risk

Siemens Medical Solutions USA, Inc recalls Artis zee systems with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for si…

Recall date
May 25, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2526-2018
FDA classification
Class II
Brand / firm
Siemens Medical Solutions USA, Inc
Sold / distributed
HI , IL, MI, NC, NY, VA, WI

Why it was recalled

Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Artis zee systems with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.