Medical device recalls Moderate risk

Siemens Medical Solutions USA, Inc recalls AXIOM Artis dFA, Model Number 7555373

Recall date
June 12, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2024-2019
FDA classification
Class II
Brand / firm
Siemens Medical Solutions USA, Inc
Sold / distributed
US Nationwide Distribution

Why it was recalled

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

AXIOM Artis dFA, Model Number 7555373

Get recall alerts

Free email alert whenever Siemens Medical Solutions USA, Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Medical Solutions USA, Inc