Siemens Medical Solutions USA, Inc recalls AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Mult…
- Recall date
- August 6, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0298-2019
- FDA classification
- Class II
- Brand / firm
- Siemens Medical Solutions USA, Inc
- Sold / distributed
- The products were distributed to the following US states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, and WY.
Why it was recalled
In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Systems radiographic exposures of the entire body.
Get recall alerts
Free email alert whenever Siemens Medical Solutions USA, Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Siemens Medical Solutions USA, Inc