Medical device recalls Moderate risk

Siemens Medical Solutions USA, Inc recalls Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194

Recall date
April 17, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2128-2020
FDA classification
Class II
Brand / firm
Siemens Medical Solutions USA, Inc
Sold / distributed
US Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.

Why it was recalled

The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194

Get recall alerts

Free email alert whenever Siemens Medical Solutions USA, Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Medical Solutions USA, Inc