Medical device recalls Moderate risk

Siemens Medical Solutions USA, Inc recalls COHERENCE Oncologist, Model Nos. 07333680 & 07351898

Recall date
January 16, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1346-2020
FDA classification
Class II
Brand / firm
Siemens Medical Solutions USA, Inc
Sold / distributed
Domestic distribution to CA, FL, GA, IA, MA, MI, NE, NJ, NV, NY, OH, OK, PA, SD, TN, WV. Worldwide foreign distribution.

Why it was recalled

A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.