Siemens Medical Solutions USA, Inc recalls Multix Fusion system is a radiographic system: Models: Multix Fusion Analog (model number 10746665) with Tube stand 3D…
- Recall date
- August 3, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0297-2019
- FDA classification
- Class II
- Brand / firm
- Siemens Medical Solutions USA, Inc
- Sold / distributed
- US nationwide distribution
Why it was recalled
Steel cables inside the lifting column of the overhead tube, may rapture without triggering the safety lock, leading to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Multix Fusion system is a radiographic system: Models: Multix Fusion Analog (model number 10746665) with Tube stand 3D V (model number 7042042 or 7042059); Multix Fusion Digital Portable (model number 10746666) with Tube stand 3D V (model number 7042042); Multix Fusion Digital Integrated (model number 10746667) with Tube stand 3D V (model number 7042042); Multix Fusion Digital Wireless (model number 10893300) with Tube stand 3D V (model number 7042042)