Siemens Symbia S Series system recalled over fire hazard
- Recall date
- July 31, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Siemens Medical Solutions USA, Inc. recalls Siemens Symbia S Series system, a diagnostic analog or digital detector based planar (2- dimensional) nuclear medicine…
- Recall number
- Z-2435-2015
- FDA classification
- Class II
- Brand / firm
- Siemens Medical Solutions USA, Inc.
- Sold / distributed
- Worldwide Distribution - USA including AL, AK, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and Internationally to Algeria…
Why it was recalled
Fuses with the improper amperage rating were installed in the Symbia S, Symbia T Series and Symbia lntevo systems. These fuses, in conjunction with a specific failure mode of the Power Controller assembly, could lead to the overheating of the assembly.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Siemens Symbia S Series system, a diagnostic analog or digital detector based planar (2- dimensional) nuclear medicine (NM) imaging system installed and used in only one location within an imaging facility or in a mobile/ transportable imaging environment.