Medical device recalls Moderate risk

Siemens Medical Solutions USA, Inc recalls SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (fu…

Recall date
April 13, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1670-2015
FDA classification
Class II
Brand / firm
Siemens Medical Solutions USA, Inc
Sold / distributed
Nationwide Distribution.

Why it was recalled

Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: Querying and retrieving patient history information and/or previous diagnosis and images from other modalities, including X-ray examinations of the urogenital area, Ultrasound examinations, Endourological interventions, Percutaneous interventions, Laparoscopy, Application of fistula, Simple procedures, Extracorporeal shock wave lithotripsy, Uroflow/urodynamics, Pediatric radiological and therapeutic applications.