Medical device recalls Moderate risk

SOMATOM Emotion 6 recalled over injury risk

Recall date
April 20, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Siemens Medical Solutions USA, Inc recalls SOMATOM Emotion 6; the intended use of computed tomography is to produce cross-sectional images of the body by computer…
Recall number
Z-1644-2015
FDA classification
Class II
Brand / firm
Siemens Medical Solutions USA, Inc
Sold / distributed
Nationwide Distribution including Puerto Rico.

Why it was recalled

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

SOMATOM Emotion 6; the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Get recall alerts

Free email alert whenever Siemens Medical Solutions USA, Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Medical Solutions USA, Inc