Siemens Medical Solutions USA, Inc recalls SOMATOM Go.Now computed tomography (CT) x-ray system table, Model Number: 11061618 Product Usage: Intended to generate…

Recall date
October 2, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0714-2018
FDA classification
Class II
Brand / firm
Siemens Medical Solutions USA, Inc
Sold / distributed
US Nationwide Distribution in the states of FL, MO, NY, TN

Why it was recalled

The table may become unstable if the nuts on the bolts were not adequately tightened and thus have the potential to loosen and several bolts may gradually loosen at the same time.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

SOMATOM Go.Now computed tomography (CT) x-ray system table, Model Number: 11061618 Product Usage: Intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by these systems can be used by a trained physician as an aid in diagnosis. These CT systems can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.