Siemens Medical Solutions USA, Inc recalls SOMATOM Perspective Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmiss…
- Recall date
- February 15, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1447-2018
- FDA classification
- Class II
- Brand / firm
- Siemens Medical Solutions USA, Inc
- Sold / distributed
- USA (nationally) Distribution.
Why it was recalled
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
SOMATOM Perspective Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.