Siemens Medical Solutions USA, Inc recalls Syngo Dynamics; Kinetdx Picture Archiving and Communications System
- Recall date
- April 14, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1601-2016
- FDA classification
- Class II
- Brand / firm
- Siemens Medical Solutions USA, Inc
- Sold / distributed
- Nationwide: AZ, CA, IL, IN, MA, MD, MI, MN, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI
Why it was recalled
Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Syngo Dynamics; Kinetdx Picture Archiving and Communications System