Siemens Medical Solutions USA, Inc recalls Syngo Imaging XS is a Picture Archiving and Communication System (PACS)
- Recall date
- November 17, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0550-2016
- FDA classification
- Class II
- Brand / firm
- Siemens Medical Solutions USA, Inc
- Sold / distributed
- AL ; CA ; CO ; CT ; FL ; GA ; IA ; IL ; IN ; KS ; MD ; MI ; MO ; NC ; NE ; NJ ; NV ; NY ; OH ; PA ; TN ; TX ; WI ;
Why it was recalled
For the Syngo Workflow SLR System with Software Ver: VB10C: Printouts may be printed in incorrect anatomical size when using syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo Imaging XS printer configuration file.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Syngo Imaging XS is a Picture Archiving and Communication System (PACS)