Siemens Medical Solutions USA, Inc recalls Uroskop Omnia max, Model No. 10762473 - Product Usage: Uroskop Omnia Max is a device intended to visualize anatomical s…
- Recall date
- January 8, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1333-2020
- FDA classification
- Class II
- Brand / firm
- Siemens Medical Solutions USA, Inc
- Sold / distributed
- US Nationwide distribution.
Why it was recalled
Siemens became aware of an issue of incorrect use of the bypass key overriding the built-in collision control creating risk of collision with obstacles or persons when the operator uses the override function.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Uroskop Omnia max, Model No. 10762473 - Product Usage: Uroskop Omnia Max is a device intended to visualize anatomical structures by converting an Xray pattern into a visible image. The system is designed primarily for urological diagnosis and the support of urological therapeutic applications such as examinations and small interventions of the urogenital tract.