Medical device recalls Moderate risk

Silex Medical, LLC recalls Signature Laparoscopic Instruments, Bullet Tip Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Con…

Recall date
January 9, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1897-2025
FDA classification
Class II
Brand / firm
Silex Medical, LLC
Sold / distributed
US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.

Why it was recalled

Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Signature Laparoscopic Instruments, Bullet Tip Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5863030.