SIN SISTEMA DE IMPLANTE NACIONAL S.A recalls Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N
- Recall date
- September 1, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0920-2025
- FDA classification
- Class II
- Brand / firm
- SIN SISTEMA DE IMPLANTE NACIONAL S.A
- Sold / distributed
- US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
Why it was recalled
Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N
Get recall alerts
Free email alert whenever SIN SISTEMA DE IMPLANTE NACIONAL S.A has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: SIN SISTEMA DE IMPLANTE NACIONAL S.A