Medical device recalls Moderate risk

Smith & Nephew, Inc. recalls LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Siz…

Recall date
December 16, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0878-2020
FDA classification
Class II
Brand / firm
Smith & Nephew, Inc.
Sold / distributed
USA: IA, NC, PA, TN. Int'l: CH, BE, GB, AE, CN, DE, NL, BR, FR.

Why it was recalled

Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).