Smith & Nephew, Inc. recalls R3(TM) 22 MM I.D., 58 MM O.D. CONSTRAINED LINER, REF 71339158, QTY: (1), STERILE EO. Product Usage: The R3 Constrained…
- Recall date
- May 6, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1737-2015
- FDA classification
- Class II
- Brand / firm
- Smith & Nephew, Inc.
- Sold / distributed
- Worldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of Australia, Switzerland, Chile, Canada, and Spain
Why it was recalled
The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
R3(TM) 22 MM I.D., 58 MM O.D. CONSTRAINED LINER, REF 71339158, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.