Medical device recalls Moderate risk

Smith & Nephew, Inc. recalls smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.

Recall date
May 2, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1921-2018
FDA classification
Class II
Brand / firm
Smith & Nephew, Inc.
Sold / distributed
US nationwide distribution.

Why it was recalled

Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.