Medical device recalls Moderate risk

Smith & Nephew, Inc. recalls Smith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WIT…

Recall date
November 9, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0573-2016
FDA classification
Class II
Brand / firm
Smith & Nephew, Inc.
Sold / distributed
US Distribution to states of:FL, MO, IN and TX.

Why it was recalled

One lot of 35 mm patella implants is labeled as 32 mm. These implants have pegs which are spaced proportionally; therefore the mismatch prevents the pegs from properly aligning with pre-drilled holes during implantation.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Smith & Nephew GENESIS(R) II/LEGION(TM) PATELLAR COMPONENT, RESURFACING ROUND, 32MM STANDARD, FOR CEMENTED USE ONLY WITH JOURNEY PEG, STERILE EO, Rx only, REF Number: 71934490 knee implant.