Medical device recalls Moderate risk

Smith & Nephew, Inc. recalls smith&nephew POLARSTEM(TM) Femoral Stems with Ti/HA (Standard, Lateral, Valgus and Collar), REF 76102260, NON STERILE,…

Recall date
July 24, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3100-2017
FDA classification
Class II
Brand / firm
Smith & Nephew, Inc.
Sold / distributed
Worldwide Distribution - US (Nationwide) and Internationally to Australia, Belgium, Norway, and Switzerland.

Why it was recalled

The Cardan joint of the Polarstem Cardan could potentially fracture or functionally fail after multiple procedures.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

smith&nephew POLARSTEM(TM) Femoral Stems with Ti/HA (Standard, Lateral, Valgus and Collar), REF 76102260, NON STERILE, QTY: (1), Rx only The subject cardan is used in conjunction with a curved stem impactor used to implant the POLARSTEM hip stem. The hip stem is mounted on the screw at the tip of the instrument, and the stem is then impacted into the femur. After impaction, the surgeon uses a knob at the end of the cardan to unscrew the implanted stem from the instrument