Smiths Medical ASD, Inc. recalls CADD-Solis VIP Ambulatory Infusion Pump; Model/Number: 21-2120-0100-02 21-2120-0100-03 21-2120-0100-06 21-2120-0100-232…
- Recall date
- April 10, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1668-2025
- FDA classification
- Class I
- Brand / firm
- Smiths Medical ASD, Inc.
- Sold / distributed
- Worldwide - US nationwide distribution.
Why it was recalled
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
CADD-Solis VIP Ambulatory Infusion Pump; Model/Number: 21-2120-0100-02 21-2120-0100-03 21-2120-0100-06 21-2120-0100-232 21-2120-0100-233 21-2120-0100-50 21-2120-0100-51 21-2120-0100-95 21-2120-0102-02 21-2120-0102-03 21-2120-0102-06 21-2120-0102-07 21-2120-0102-08 21-2120-0102-12 21-2120-0102-13 21-2120-0102-14 21-2120-0102-15 21-2120-0102-17 21-2120-0102-237 21-2120-0102-238 21-2120-0102-247 21-2120-0102-249 21-2120-0102-51 21-2120-0102-78 21-2120-0102-92 21-2120-0102-97 21-2120-0102-98 21-2120-0103-01 21-2120-0103-02 21-2120-0103-03 21-2120-0103-06 21-2120-0103-07 21-2120-0103-08 21-2120-0103-12 21-2120-0103-14 21-2120-0103-15 21-2120-0103-17 21-2120-0103-249 21-2120-0103-51 21-2120-0103-78 21-2120-0104-01 21-2120-0104-231 21-2120-0105-00 21-2120-0105-01 21-2120-0105-02L 21-2120-0105-03L 21-2120-0105-06L 21-2120-0105-07L 21-2120-0105-08 21-2120-0105-08L 21-2120-0105-12L 21-2120-0105-13L 21-2120-0105-14L 21-2120-0105-15L 21-2120-0105-17L 21-2120-0105-50 21-2120-0105-50L 21-2125-0104-01 21-2125-0105-03L 21-2125-0105-07L 21-2125-0105-08L 21-2125-0105-50L 21-2127-0104-01 21-2127-0105-01 21-2127-0105-02L 21-2127-0105-03L 21-2127-0105-06L 21-2127-0105-08L 21-2127-0105-50L
Get recall alerts
Free email alert whenever Smiths Medical ASD, Inc. has a new recall — straight from official government data. Unsubscribe anytime.