Medical device recalls Moderate risk

Smiths Medical ASD Inc. recalls Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced durin…

Recall date
December 3, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0653-2022
FDA classification
Class II
Brand / firm
Smiths Medical ASD Inc.
Sold / distributed
Distribution US nationwide, including PR, VI, GU, and MP. There was government/military/foreign distribution.

Why it was recalled

There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced using replacement main board part numbers G6000361, G6000435, G6001260, G6001560, G6001562 and G6002728.

Get recall alerts

Free email alert whenever Smiths Medical ASD Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Smiths Medical ASD Inc.